F0 Value in Steam Sterilization - Pharma GxP (2024)

Sterilization means the process of reducing the microbial population to the acceptable levels and maintain the sterility of the drugs or medical devices.

As appropriate, there are different types of sterilization techniques used to make contamination-free product contact parts. For example, Steam Sterilization, Dry Heat Sterilization, Ethylene Oxide Sterilization, etc.

Out of them, the F0 value (read as F Zero) is designed for moist heat sterilization (or steam sterilization).

In this article, you’ll learn:

  • Mechanism of steam sterilization
  • Background to and Calculation of the F0 value
  • Understanding the meaning of the resulted F0 value
  • Practical examples of the F0 value

So if you want to excel in the F0 calculations, this guide is for you.

First thing first, let us see the fundamentals of moist heat kinetics through mathematical and biological angles. Later, we’ll see how it connects with the F0 value concept.

Table of Content

Steam Sterilization Kinetics (Math vs. Bio)

Assume a microbial species that has contaminated the system, is in contact with saturated steam at a constant temperature.

Mathematically, sterilization develops as a first-order chemical reaction (more specifically, a biochemical reaction).

The rate of the reaction then becomes:

F0 Value in Steam Sterilization - Pharma GxP (1)

Where,
N – Number of microbes present in the system
t – exposure time to the microbes with steam
K – the rate of reaction’s constant (species dependent)

Re-arranging to integrate,

F0 Value in Steam Sterilization - Pharma GxP (2)

Now integrating,

F0 Value in Steam Sterilization - Pharma GxP (3)

Converting and simplifying the logarithmic value to exponential,

F0 Value in Steam Sterilization - Pharma GxP (4)

where,
N0 – no. of microbes initially
t – steam exposure time (sterilization hold time)
N – no. of microbes at the end of sterilization hold
k – the rate of reaction constant (microbes dependent); now this k = K / 2.303. Here, K is called the Death Rate constant.

The equation (1) hence shows that no. of microbes decreases exponentially with increasing sterilization hold time. Also, the time required to decrease the microbial amount to a pre-defined value depends on its initial number.

What is D-Value (Decimal Decay Time)?

D-value is the time (t) required at a specified temperature (T) to reduce the microbial population from 100% to 10% (i.e. 90% or 1 log reduction).

Definition of D-value

Mathematically, the D-value is the reciprocal of rate of reaction constant (k) and therefore;

K = t -1

By putting this value in earlier equation (1) we get,

N = 0.1*(N0)

The D-value plays an important role in determining hold time for the system under the sterilization.

More commonly, D-value is considered as “1” if there is no precise experimental data available. A typical range for D-value is 0 to 2 min. at around 121°C.

In other words, D-value is the time required in minutes to destroy the 90% population of the micro-organisms.

D-Value vs. Time

Let us understand the importance of D-value with an example where we want to reduce micro-organisms with moist heat sterilization.

  • Initial no. of microbes = 100 i.e. 102
  • Target reduction of microbes = 10-6
  • Constant Temperature of sterilization process (T) = 121°C

The table below is an experimental data explaining how the D-value affects microbial reduction time:

D-Value
(min.)
Time Required for Microbial Reduction
(min)
Initial No. of Micro-organismsFinal No. of Micro-organisms
0.5410010-6
1810010-6
21610010-6

This data means, at 121°C, the residual contamination of 10-6 is gained from the initial contamination of 102 in:

  • 8 mins if D121=1,
  • 4 mins if D121=0.5 and
  • 16 mins if D121=2.

Another way to calculate the D-value is by using the following formula.

D = t / (log Co – log C)

where, t – exposure time in minutes
Co – initial number of micro-organisms
C – final number of micro-organisms

Please remember, the log values we’re discussing here are log10 and not loge.

European Pharmacopoeia [E 10.0, 5.5.1] have defined Sterility as Sterility Assurance Level (SAL) = 10-6.

Theoretically equivalent to SAL, United States Pharmacopoeia [USP 43, <1229>] have defined a must for sterility as Probability of Non-Sterile Unit (PNSU) = 10-6 i.e. the probability of identifying a non-sterile unit or micro-organism in a batch or lot.

In the above table, the SAL values are 100 and 10-6. Hence, the probability of identifying non-sterile microbe is one in 100 and one in a million, respectively.

What is z-Value (Temperature Coefficient)?

Practically speaking, the temperature wouldn’t be that flawless throughout the exposure. Rather it would be dynamic with time resulting in changing D-value.

The z-value is used to correlate the heat resistance of the micro-organisms to the changing temperature.

z-value is the number of degrees the temperature is required to be increased which will cause a 10-fold variation in the D-value.

Definition of Z-Value

The z-value is considered as 10°C for the temperature range from 100 to 130°C for steam sterilization. Generally, bacteria that form spores have a z-value range of 10 to 15°C while the non-spore-forming ones have a range of 4 to 6°C.

Approximately D-value varies by 10 times for variation of 10°C. This doesn’t mean that D-value varies by 1 time for 1°C variation. The rationale for this is beyond the scope of this topic. To keep things simple, only note that “Variation in 1°C requires a variation of D-value by 25-30%”.

The effect of changing temperature decreases considerably upon:

1. Changing the sterilization method
2. Raising the system temperature

For example, for dry heat sterilization of around 200°C, the z-value is considered as 20°C instead of 10°C.

Even a small temperature incline will impact steam sterilization. At the same time, it may be insignificant in dry heat sterilization.

At 121.1°C, practically no microbe has precise D=1 minute and z=10°C but co-firing these two parameters to estimate the F0 value provides sufficient room for error.

The D-value is measured in minutes while the z-value is a temperature coefficient measured in °C.

Let us see, how the D-value and z-value help to reach a meaningful outcome for the sterilization effectiveness.

F0 Value (Equivalent Exposure Time)

F0 value is the equivalent exposure time at 121.1°C to that of the actual exposure time at a variable temperature calculated with a temperature coefficient of the destruction of 10°C.

F0 Value Definition
F0 Value in Steam Sterilization - Pharma GxP (5)

where,
Δt – the time interval between two temperature readings
T – the temperature at time t of the product under sterilization
z – temperature coefficient (assumed as 10°C)

The term next to Δt in equation (2) is called Lethality Rate. You can calculate it individually or by averaging complete sterile hold data, the result will be equivalent.

Testing F0 Value And Its Meaning

Consider sterilization hold time of 30 min. at constant 121.1°C.

Put this in equation (2),

F0 Value in Steam Sterilization - Pharma GxP (6)

Solving we get, F0 = 30 min. (This is an ideal condition for steam sterilization)

But what if the temperature is increased or decreased from 121.1°C (ideal condition)?

Let’s find out…

Example 1 (Below Ideal Condition)

Consider sterilization hold time of 30 minutes at constant 110°C instead of 121.1°C, in a similar manner as above,

F0 Value in Steam Sterilization - Pharma GxP (7)

Solving we get, F0=2.33 min.

Hence, 30 minutes of sterilization at 110°C is lethally equivalent to 2.33 min. of sterilization at 121.1°C. (Compare this sentence with the definition of F0 value for easy understanding)

The crux of the example is,
121.1°C 30 min. (Expected condition)
110°C 2.33 min. (Actual condition compared to the expected condition)

Compared to the expected condition, the effective sterilization cycle time was only 2.33 instead of 30 minutes because of the lowered temperature of 110°C.

Now, let us understand “how long will it take to complete the sterilization?” for this condition. Meaning, if we maintain temperature 110°C constantly, how much hold time is required to achieve an F0 value of 30 min.

By reverse calculation, we can do this.

Let’s fix the F0 value to 30 minutes. The unknown part then would be Δt (sterile hold time).

Data we have now is F0=30 minutes and Temperature T=110°C

Putting these values in equation (2),

F0 Value in Steam Sterilization - Pharma GxP (8)

Solving we get, Δt=386 min.

When the temperature of the system under sterilization is maintained at a constant 110°C, the time of 386 min. (i.e. Sterile Hold Time) would be required to achieve the lethal effect of 121.1°C at 30 min.

Example 2 (Above Ideal Condition)

Consider sterilization hold time of 30 minutes at constant 125°C instead of 121.1°C, in a similar manner as above,

F0 Value in Steam Sterilization - Pharma GxP (9)

Solving we get, F0=73.64 min.

Hence, 30 minutes of sterilization at 125°C is lethally equivalent to 73.64 min. of sterilization at 121.1°C. (Compare this sentence with the definition of F0 value for easy understanding)

Simply put,
121.1°C 30 min. (Expected condition)
125°C 73.64 min. (Actual condition compared to the expected condition)

Compared to the expected condition, the effective sterilization cycle time was 73.64 instead of 30 minutes because of the elevated temperature of 125°C.

Let us understand “how long will it take to complete the sterilization?” for this condition. Meaning, if we maintain temperature 125°C constantly, how much hold time is required to achieve an F0 value of 30 minutes.

By reverse calculation, we can do this. Let’s fix the F0 value to 30 min. The unknown part then would be Δt.

Data we have now is F0=30 minutes and Temperature T=125°C

Putting these values in equation (2),

F0 Value in Steam Sterilization - Pharma GxP (10)

Solving we get, Δt = 12.24 min.

When the temperature of the system under sterilization is maintained at a constant 125°C, the time of 12.24 min (i.e. Sterile Hold Time) would be required to achieve the lethal effect of 121.1°C at 30 min.

You may use this F0 Value Calculator to explore the different steam sterilization scenarios.

Key Takeaways

F0 value provides lethal equivalence between expected and practical conditions. When noting down the values of lethal doses, D-value and z-value (if obtained) become very important to evaluate the precise F0 value.

Sterilization cycles can fail even if the sterile hold time is completed if the average temperature of the system is below 121.1°C. This is because residual microorganisms cannot be reduced completely.

When the average temperature of the system is above the set value of 121.1°C, the sterilization may be lethally equivalent even before the sterile hold time completes.

Note: This lethal equivalence has technical value only when the steam used for sterilization is Saturated. Otherwise, it is pointless. To know more, read out exact steam quality requirements.

F0 value calculation can be done either post sterilization or in real-time depending upon the chosen method.

The ‘post sterilization’ approach helps in analyzing and evaluating the lethal equivalence once the cycle has finished. While real-time calculation with a suitable controller help actually control the sterilization cycle till it achieves the preset F0 value.

Most of the pharmaceutical companies practicing steam sterilization rely on the calculation of F0 value for analysis purposes only. Though the logic-based controllers are capable of effectively driving the sterilization cycles based on the F0 value.

Experts recommend using the F0 value when the steam sterilization at 121°C for 15 min. is ineffective. In fact, Maria Luisa Bernuzzi, a technical center expert at Fedegari group explains this so simply…

F0 Value in Steam Sterilization - Pharma GxP (11)

Indeed factual! Further, she also explains the overkill approach as per EU regulations…

F0 Value in Steam Sterilization - Pharma GxP (12)

This means the steam sterilization cycle can be F0-based or time-based but the important thing is, whichever way you go, you should meet the required lethality.

If you want to read about the validation approach for steam sterilization or autoclave, you can read it here.

F0 Value: FAQs

What is f0 value?

F0 value is an equivalent exposure time at 121.1°C to that of the actual exposure time at a different temperature calculated with a temperature coefficient z-value of 10°C. Read more in the above article.

How do you calculate f0 value in autoclave?

To calculate the F0 value, average the temperature of each probe during the sterile hold and then average the different temperatures to get one single temperature. Use the formula shown in the above article to estimate the F0 value.

What is D value and Z value?

D-value is the time (t) required at a specified temperature (T) to reduce the microbial population from 100% to 10% (1 log reduction). z-value is the number of degrees the temperature required to increase, which will effect a 10-fold variation of D-value.

Now, turning it over to you.

How do you interpret the F0 value? Is it the same as this? Or maybe something else.

Either way, comment below.

F0 Value in Steam Sterilization - Pharma GxP (2024)

FAQs

What is the F0 value of steam sterilization? ›

The term “F0 ” is defined as the number of equivalent minutes of steam sterilization at 250°F (121°C) delivered to a load (product)1.

What is F0 for autoclave? ›

The vast majority of loads sterilized in the laboratory setting are sterilized at 121.1°C/250°F for 30-90 minutes. F0 is the accumulated lethality in minutes at 121.1°C/250°F.

What is the importance of F value in sterilization? ›

The F value\**\ for a process is the number of minutes required to kill a known population of microorganisms in a given food under specified conditions. This F value is usually set at 12 D values to give a theoretical 12 log cycle reduction of the most heat-resistant species of mesophilic spores in a can of food.

How do you find F0? ›

There are also different ways to evaluate the F0 value. One way is to average the individual lethal doses over a sterile hold and then test the F0 value. The second way is to average the temperature and then calculate average lethal doses.

What is the F0 value guideline? ›

The F0 value is defined as the thermal lethality time required to eliminate all microorganisms present in foods, by exposing them to a temperature of 121.1ºC and it is expressed in minutes.

What does F0 mean sterilization? ›

"Fo means the equivalent amount of time, in minutes at 121°C or 250 F, which has been delivered to a product by the sterilization process". For the practical calculation of F0, "a z-value of 10°C or 18 F is assumed; the term z-value.

What is the F0 value for pasteurization? ›

Therefore, a value of F0 = 5.5 minutes is equivalent to 5.5 sterilisation units (At 121.11°C), and 20 pasteurisation units are equivalent to exposing the product for 20 minutes at the reference temperature (normally 60°C).

What is the value of F0 in UHT? ›

F0 value. F0 = 1 minute after the product is heated at 121.1 °C for one minute. Based on that F0 value 1 minute gives 4 log reduction of Clostridium botulinum spores. To obtain commercially sterile milk from good quality raw milk in practice, UHT plants are designed to achieve a minimum F0 value of 5 – 6 minutes.

How do you confirm autoclave sterilization? ›

There are three indicators that may be used to detect the efficacy of the autoclave process: (1) Physical: pressure and temperature recording devices, (2) Chemical: indicators that change color after being exposed to specific temperatures, such as temperature sensitive tape.

How much F value is significant? ›

You perform the F test by looking for the appropriate p-value in the computer analysis and interpreting the resulting significance level, as we did in Chapter 10. If the p-value is more than 0.05, then the result is not significant. If the p-value is less than 0.05, then the result is significant.

Why are F values always positive? ›

The F-distribution cannot take negative values, because it is a ratio of variances and variances are always non-negative numbers. The distribution represents the ratio between the variance between groups and the variance within groups.

What is the lethal rate F value? ›

It is the measure of the death rate of an organism. For example if F250 = 3.0, this is the time in minutes at 250°F (121° C) required to achieve sterility of a suspension containing a known number of cells or spores (commercial sterility).

What is the F0 range? ›

Typically fundamental frequencies lie roughly in the range 80 to 450 Hz, where males have lower voices than females and children. The F0 of an individual speaker depends primarily on the length of the vocal folds, which is in turn correlated with overall body size.

How to calculate F0 in sterilization? ›

As i saw in a document, Fo is calculated by considering D value in equations. For example the equation is written as Fo=1/D*(10^T-121.11/z)*∆t.

What does F0 measure? ›

Pitch corresponds to the physiological parameter of the frequency of vibration of the vocal folds (aka vocal cords) which can be measured in cycles per second (cps), or the equivalent acoustic parameter of fundamental frequency (f0), measured in hertz (Hz).

What is an fo value? ›

The Fo value is a concept associated with the thermal destruction of microorganisms relative to food preservation and canning. The F value is used to designate the time necessary to destroy a given number of microorganisms at a reference temperature, usually 121°C for spores or 60°C for vegetative cells.

What is the quality of steam for sterilization? ›

Steam systems should be designed to ensure that the steam delivered to the sterilizer is saturated steam having a steam quality of 97% to 100%. Steam of poor quality can contribute to wet packs and to sub-optimal steam sterilization cycles that might or might not be identified by biological monitoring.

Top Articles
Latest Posts
Article information

Author: Mrs. Angelic Larkin

Last Updated:

Views: 5861

Rating: 4.7 / 5 (47 voted)

Reviews: 94% of readers found this page helpful

Author information

Name: Mrs. Angelic Larkin

Birthday: 1992-06-28

Address: Apt. 413 8275 Mueller Overpass, South Magnolia, IA 99527-6023

Phone: +6824704719725

Job: District Real-Estate Facilitator

Hobby: Letterboxing, Vacation, Poi, Homebrewing, Mountain biking, Slacklining, Cabaret

Introduction: My name is Mrs. Angelic Larkin, I am a cute, charming, funny, determined, inexpensive, joyous, cheerful person who loves writing and wants to share my knowledge and understanding with you.